medical hospital infant baby incubator



Brief Introduction:

Air mode control, set temperature and air temperature are displayed separately;
Natural air flow humidity
Failure alarm indication;
RS-232 connector
YD-YP100B are configured with fixed Neonate Bilirubin Phototherapy Equipment model: YD-XHZ-90 (with blue fluorescence lamps)
DC motor
 
Standard Configuration:
YD-YP100A: Based on YD-YP100 ,added with the side door
YD-YP100B: Based on YD-YP100A, added with the Neonate bilirubin phototherapy equipment (fixed model)
 
Optional Configuration:
>37°C temperature override mode fundation, Baby mode, Configured with fixed neonate bilirubin phototherapy equipment model: YDXHZ-90L (with LED blue light) and SPO2
 
Specification:
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 850VA
Mode of control: Air mode control by micro-computer
Air temperature control range: 25°C -37°C (override mode: 37°C -39°C is optional)
The temperature difference between incubator temperature and average incubator temperature(under steady temperature condition)≤0.5°C
Temperature uniformity≤0.8°C
Noise inside hood: 45dB(A)
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on.
 
Environmental temperature:
Operating range: 20 °C -30°C (Set temperature should be higher 3°C than the environment temperature)
Ambient air movement rate: < 0.3m/s

Package
Package each unit is packed into one carton
Dimension 1085mm x 640mm x 1075mm
Gross Weight 83kg